For US healthcare professionalsImportant Safety InformationReport a product issue(404) 594-4280

Performance data

Directional Cryoprobe isotherm data.

Isotherm dimensions for both modes from laboratory testing, as stated in the device labeling.

CIRCUMFERENTIAL · 360°10 mm
0 °C−20 °C−40 °C
Circumferential (360°), to scale.
DIRECTIONAL · ~180°COLD SIDE10 mm
0 °C−20 °C−40 °C
Directional (~180°), to scale, on the cold side of the probe.
Circumferential (360°)
IsothermWidth (mm)Length (mm)
0 °C3773
−20 °C2362
−40 °C1744

Protocol: 10-minute freeze (100% duty cycle), 5-minute passive thaw, 10-minute freeze · Accuracy ±0.9 mm width · ±2.8 mm length

Directional (~180°)
IsothermWidth (mm)Length (mm)
0 °C1149
−20 °C630
−40 °C311

Protocol: 12-minute built-in directional protocol · Accuracy ±1.8 mm width · ±6.6 mm length

Isotherm data collected in 37 °C ± 2 °C ultrasound gel in a water bath.