Performance data
Directional Cryoprobe isotherm data.
Isotherm dimensions for both modes from laboratory testing, as stated in the device labeling.
0 °C−20 °C−40 °C
0 °C−20 °C−40 °C
| Isotherm | Width (mm) | Length (mm) |
|---|---|---|
| 0 °C | 37 | 73 |
| −20 °C | 23 | 62 |
| −40 °C | 17 | 44 |
Protocol: 10-minute freeze (100% duty cycle), 5-minute passive thaw, 10-minute freeze · Accuracy ±0.9 mm width · ±2.8 mm length
| Isotherm | Width (mm) | Length (mm) |
|---|---|---|
| 0 °C | 11 | 49 |
| −20 °C | 6 | 30 |
| −40 °C | 3 | 11 |
Protocol: 12-minute built-in directional protocol · Accuracy ±1.8 mm width · ±6.6 mm length
Isotherm data collected in 37 °C ± 2 °C ultrasound gel in a water bath.
See the system in person.
Request a demonstration, labeling, or a conversation with our team.